BC-120H
DERUK
V-2600200
Digital Electronic Colposcope is widely applied in gynecological routine examination, cervical lesion screening and follow-up observation.
This cost-effective model integrates a 0.12MP color CCD camera to capture clear real-time images of cervical tissue.
Adjustable magnification and excellent color restoration help clinicians identify subtle surface lesions effectively.
Equipped with dimmable LED light source, the device offers soft and uniform illumination without glare.
The multi-language system facilitates operation for global users.
Ergonomic design enables easy daily operation, while stable long-term working performance lowers maintenance costs.
It is widely favored by primary medical centers, outpatient clinics and medical examination centers for daily cervical screening work.
No | Technical Specifications |
1 | Complies with the electrical safety requirements of GB 9706.1-2007. |
2 | Conforms to the electromagnetic compatibility requirements of YY 0505-2012. |
3 | Under specific magnification, the lens working distance shall cover no less than 200 mm to 300 mm. |
5 | Working distance ≥ 100 mm at 3× magnification. |
6 | Working distance ≥ 15 mm at 18× magnification. |
7 | Depth of field ≥ 120 mm at 6× magnification; depth of field ≥ 5 mm at 18× magnification. |
8 | Optical zoom: 1–36×, linear correlation coefficient R of magnification ≥ 0.95; electronic magnification up to 280×. |
9 | Illuminance is adjustable. Under maximum illuminance, the output beam illuminance of the instrument shall be ≥ 1600 Lx within the working distance of 200–300 mm. |
10 | White light color temperature: 4000 K ~ 5500 K. |
11 | The diameter of illuminated spot ≥ 60 mm at working distance of 200–300 mm. |
12 | Illuminance uniformity ≤ 110%. |
13 | Temperature rise at the center of irradiated surface ≤ 11 °C when working distance L = 200 mm. |
14 | System resolution ≥ 550 TVL, spatial resolution ≥ 10 lp/mm, image resolution ≥1024*1200 |
15 | Maximum color rendering error ≤ 30 NBS, average color rendering error ≤ 20 NBS; average camera saturation value: 95% ~ 120%. |
16 | Image geometric distortion ≤ 3%. |
17 | 2-megapixel color CCD camera. |
Digital Electronic Colposcope Real Clinical Shot of Cervical Lesion Screening